Recall Z-2556-2025 ยท OVD Intervertebral Fusion Device With Integrated Fixation, Lumbar

Omnia Medical

TiBrid Stand Alone Intervertebral Body Fusion Device, SA Angled Punch Awl, Sizes: 30 degree, 45 degree and 60 degree.

DateSep 11, 2025 (initiated Jul 15, 2024)
StatusOpen, Classified
ReasonFailure of fusion system instruments in the field.
Root causeDevice Design
ActionConsignees were called on or around 8/1/2024 to notify them of this recall of instruments. Affected instruments were to be sent back to Omnia Medical and replaced with instruments that underwent rework in the form of design changes aimed at increasing their durability.
Quantity15 units
DistributionUS Nationwide distribution in the states of AZ, CA, NJ, NV, OK & WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K203207Omnia Medical TiBrid-SA
Omnia Medical, LLC, Dec 23, 2020