Recall Z-2491-2025 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayStation with the following product descriptions: 1. RayStation 11A, Software Version: 11.0.0.951. 2. RayStation 11A Service Pack 1, Software Version: 11.0.1.29. 3. RayStation 11A Service Pack 2, Software Version: 11.0.3.116. 4. RayStation 11A Service Pack 3, Software Version: 11.0.4.15. Product Description: Radiation Therapy Treatment Planning System.

DateSep 2, 2025 (initiated Aug 8, 2025)
StatusOpen, Classified
ReasonDICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
Root causeUnder Investigation by firm
ActionRaySearch notified consignees on about 08/08/2025 via email. Consignees were informed that as temporary solution, affected objects can be re-exported via RayGateway to generate new UIDs. They were instructed to inform planning staff and all users about this workaround, inspect your product and identify all installed units with the above software version number(s), and complete and return the response form. The issue will be resolved in the next version of RayStation, scheduled for market release in December 2025 (subject to market clearance in some markets). If customers wish to continue usin
Quantity3
DistributionWorldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL, IA, KY, LA, MD, MN, MS, NY, NC, OH, OK, SC, WA, WI and the countries of Canada, China, Colombia, France, Germany, Hong Kong S.A.R., China, Italy, Japan, Netherlands, Poland, Spain, Sweden, Switzerland, ENG, Thailand, Turkey,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211867RayStation 11.0
RaySearch Laboratories AB (PUBL), Sep 8, 2021