Recall Z-2543-2025 ยท LLZ System, Image Processing, Radiological

Philips Medical Systems Nederland B.V.

IntelliSpace Cardiovascular, Software 8.0.0.4.

DateSep 9, 2025 (initiated Aug 8, 2025)
StatusOpen, Classified
ReasonSoftware issue that results in the display of outdated information.
Root causeSoftware design
ActionAn URGENT Medical Device Correction notice dated 8/7/25 was mailed to consignees. The notification includes instructions for continued safe use of devices, which includes: 1. Ensure all finding codes with variables/modifiers are populated correctly on the report screen. Do not leave any FitB Modifier Code Fields Blank. 2. Verify report statements match the selections in the work area before finalizing report. Systems may continue to be used in accordance with the intended use and per the additional instructions. The provided recall notification is to be shared with all users of the device a
Quantity4 units
DistributionUS Nationwide distribution in the states of GA, NC & TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K153022Philips IntelliSpace Cardiovascular
Philips Medical Systems Nethrlands BV, Nov 18, 2015