Recall Z-2466-2025 ยท NWU Endoscope Introducer Kit

American Contract Systems Inc

AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) CYSTO, Model/Item Number ASCY24A; 2) CYSTO PACK, Model/Item Number BUCY78E; 3) CYSTO PACK, Model/Item Number LMCP22T; 4) TUR BASIN PACK, Model/Item Number LMTU31T; 5) CYSTO, Model/Item Number SACY80S; 6) UROLOGY PACK, Model/Item Number SPUR64F.

DateAug 28, 2025 (initiated Jul 14, 2025)
StatusOpen, Classified
ReasonDuring an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Root causeUnder Investigation by firm
ActionOwens & Minor issued an "URGENT: MEDICAL DEVICE CORRECTION" notice to its consignees on 07/15/2025 via email. The notice explained the problem, potential risk to health, and requested the following: ACTIONS TO TAKE: " Discontinue use of all affected components. " Add warning labels to affected procedure trays to ensure the components are discarded. A labeling template is enclosed for printing labels which can be affixed on all affected kits in inventory. This label should be applied to each affected procedure tray and product case in a prominently visible location to end-users. The locat
Quantity1127 units
DistributionUS distribution to states of: IL and LA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.