Recall Z-2239-2025 ยท MUJ System, Planning, Radiation Therapy Treatment

Philips Medical Systems (Cleveland) Inc

Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software

DateAug 1, 2025 (initiated Jul 17, 2025)
StatusOpen, Classified
ReasonDue to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,
Root causeSoftware change control
ActionOn July 17, 2025, Philips issued a "Urgent Medical Device Correction" Notification to affected consignees. Philips asked consignees to take the following actions: 1. You may continue to use your system(s) in accordance with the intended use and by following the recommendations below. 2. To prevent the issue from occurring, when using SPR generation, always verify data sets are from same acquisition and matching reconstruction parameters. 3. Ensure that relevant clinical staff, including physicians, physicists, and dosimetrists, are informed about the potential for incorrect dataset calcula
Quantity149 Serial Numbers
DistributionWorldwide - US Nationwide distribution in the state of WV and the countries of Macao & United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170086Pinnacle Radiation Therapy Planning System
Philips Medical Systems (Cleveland), Inc., Feb 9, 2017