ClearanceBrief
Recall Z-2422-2025 ยท NDP Accessories, Implant, Dental, Endosseous

Dentsply IH, Inc.

Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04S / 6802034 Product Description: The Atlantis intraoral feature locating object for suprastructures (IO FLO-S) is intended to support the prosthetic procedure with the aim of restoring chewing function in partial or complete edentulous patients, by enabling digital impression taking on implant or abutment leve

DateAug 25, 2025 (initiated Jul 24, 2025)
StatusOpen, Classified
ReasonThe defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.
Root causeUnder Investigation by firm
ActionOn July 24, 2025, URGENT MEDICAL DEVICE RECALL letters were sent to customers. Action to be taken by the Customer: 1. Stop use of the affected lots immediately. 2. If the IO FLO-S has already been used for scanning, please immediately contact the dental laboratory that manufactured the dental restoration and direct the laboratory to contact Dentsply Sirona Atlantis Customer Service, immediately (see contact data below) to verify if the scans are correct. 3. If the IO FLO-S has already been used and a structure has been manufactured and placed into the patient, please immediately contact the p
Quantity780 units
DistributionWorldwide - US Nationwide distribution in the states of IL, LA, MA, MN, MO, ND, NJ, NM, OH, OK, PA, TX, WV and the countries of AT, AU, BE, CA, DE, DK, ES, FI, FR, IT, JP, NL, NO, PL, QA, SE, UA, UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.