ClearanceBrief
Recall Z-2332-2025 ยท FRN Pump, Infusion

Baxter Healthcare Corporation

Novum IQ LVP (Large Volume Pump), Product Code 40700BAXUS, infusion pump

DateAug 29, 2025 (initiated Aug 4, 2025)
StatusOpen, Classified
ReasonBaxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).
Root causeSoftware design (manufacturing process)
ActionBaxter issued an "URGENT MEDICAL DEVICE CORRECTION" to its consignees on 08/04/2025 via USPS first class mail. The notice described the problem, hazard involved, and actions to be Taken by customers: 1. Users should rely on their clinical judgment. When providing patient care involving high risk medications and/or critical illness, users should consider utilizing an alternate pump, if available and appropriate for the patient's treatment. 2. If an alternate pump is not available or appropriate for patient care you may continue to use the Novum IQ LVP and Novum IQ Syringe Pump consistent wit
Quantity43,922 units
DistributionUS distribution including Puerto Rico and OUS (International) Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211122Novum IQ Large Volume Pump
Baxter Healthcare Corporation, Mar 29, 2024