Recall Z-2480-2025 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

Medtronic Perfusion Systems

DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12113 LV VENT 13FR, Model Number 12113

DateSep 3, 2025 (initiated Aug 6, 2025)
StatusOpen, Classified
ReasonThe catheters may not retain their shape.
Root causeProcess change control
ActionMedtronic Cardiac Surgery issued an Urgent Medical Device Recall notice to its consignees on 08/06/2025 via UPS 2nd day delivery. The notice explained the issue with the device, potential harms if used, and requested the following: Customer Actions: Medtronic requests that you take the following actions: " Review your inventory for listed product. " Immediately identify and quarantine all unused, listed product in your inventory. " " " Please share this notification with others in your organization as appropriate. If product listed above has been forwarded to another facility, please noti
Quantity34549 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K834039INFANT VENT CATHETER 12001 W/SIDE VENT
Dlp, Inc., Jan 3, 1984