Recall Z-2347-2025 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Philips Ultrasound, Inc

3D9-3v Transducer Probe

DateAug 22, 2025 (initiated Jul 3, 2025)
StatusOpen, Classified
ReasonTo provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Root causeDevice Design
ActionAn URGENT Medical Device Correction Notice, dated 7/3/25, was mailed to consignees. This notice instructs consignees to locate all Philips Ultrasound transducers in their facility and enter their serial numbers into https://www.transducer.philips.com to obtain the manufacturing certificate for each of their devices. This certificate should be retained with its system documentation. Using the manufacturing certificate consignees are to calculate the useful life of their device. The expected useful life for Transesophageal transducers, X-Matrix transducers, and Lumify transducers is 5 years; The
Quantity171,322 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov