Recall Z-2671-2025 ยท LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation

bioMerieux, Inc.

VITEK 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. Export only

DateSep 29, 2025 (initiated Jun 26, 2025)
StatusOpen, Classified
ReasonPotential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards
Root causeUnder Investigation by firm
ActionOn 06/27/2025, the firm sent an "URGENT FIELD SAFETY NOTICE" letter to customer informing them that VITEK 2 users, who are using VITEK 2 GN-AST cards with Colistin formulation, cs02n, have reported obtaining false resistant Colistin results with Acinetobacter baumannii complex and Pseudomonas aeruginosa isolates. Customers are required to: bioMrieux recommends you confirm any resistant colistin (cs02n) results for multidrug resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex isolates via an alternate method for all existing and future AST cards with cs02n. If appropriate for
Quantity235269 cards
DistributionInternational distribution in the countries of Albania, Armenia, Australia, Austria, Belgium, Bosnia, Bulgaria, China, Croatia, Cyprus, Czech Republic/Slovakia, France, Germany, Greece, Iberia, India, Israel, Italy, Japan, Lithuania, Macedonia, Malaysia/Brunei, Moldova, Netherlands, Pakistan, Polan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.