Recall Z-2621-2025 ยท QFG Alternate Controller Enabled Insulin Infusion Pump

Tandem Diabetes Care, Inc.

t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)

DateSep 24, 2025 (initiated Jul 3, 2025)
StatusOpen, Classified
Reasonan app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.
Root causeSoftware design
ActionOn 07/03/2025, the firm sent an "URGENT FIELD SAFETY NOTICE" it distributor to notify customers that users operating an Android phone with language setting set as Hebrew may experience a blank screen on the mobile application during the initial pairing sequence. The language setting must be set to English for the pairing sequence to work successfully. However, if users set their language back to a right-to-left language after pairing, or do not encounter the initial issue, the mobile application will display discrepancies. This discrepancy may cause a user to make incorrect therapy decisions
Quantity19
DistributionInternational distribution to the country of Israel.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K232380t:slim X2 Insulin Pump with Interoperable Technology
Tandem Diabetes Care, Inc., Nov 3, 2023