Recall Z-2322-2025 ยท MSW System, Test, Thyroglobulin

Beckman Coulter, Inc.

Access Thyroglobulin assay, a simultaneous one-step immunoenzymatic assay, Catalog Number 33860

DateAug 13, 2025 (initiated Jul 10, 2025)
StatusOpen, Classified
ReasonAccess Thyroglobulin reagent lot 439163 may generate erroneously high patient results. Falsely increased Access Thyroglobulin results may lead a physician to pursue unnecessary diagnostic imaging studies and/or inappropriate therapy adjustments in patients being monitored for residual or recurrent thyroid cancer.
Root causeUnder Investigation by firm
ActionAn URGENT MEDICAL DEVICE RECALL notification dated 7/2/25 was sent to customers. ACTION: " Discontinue using Access Thyroglobulin reagent lot 439163 and discard all remaining reagent packs from this lot. " At the discretion of the medical director, a retrospective review of patient results generated with Access Thyroglobulin reagent lot 439163 should be performed, which includes reviewing patient history, to assess the need for retesting. " Please contact your local Beckman Coulter representative for replacement product requests and to receive updates on availability and delivery. RESOLUTION:
Quantity453 units
DistributionWorldwide - US Nationwide and the countries of Brazil, Canada, Georgia, Italy, Malaysia, Mexico, Philippines, Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031269ACCESS THYROGLOBULIN REAGENTS ON THE ACCESS IMMUNOASSAY SYSTEMS
Beckman Coulter, Inc., May 2, 2003