Recall Z-2437-2025 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

VANTIVE US HEALTHCARE LLC

PrisMax V2-US, Hemodialysis Delivery System, Product Codes: 955626, 955558, 955627, 955701

DateAug 26, 2025 (initiated Jul 28, 2025)
StatusOpen, Classified
ReasonLow deaeration chamber level at post priming tests may lead to PrisMax System alarm T2309: Air Detected in Prime resulting in multiple additional priming cycles
Root causeUnder Investigation by firm
ActionVantive issued an Urgent Medical Device Correction notice to it consignees on 0728/2025 via USPS first class mail. The notice explained the issues with the devices, hazard involved, and requested the following: Actions to be Taken by Customers 1. Operators may continue to safely use the PrisMax Systems and TherMax Blood Warmer Units per the associated Operator s Manuals and on-screen instructions. If further assistance is required, please contact your local Vantive account representative. Please ensure that all operators of these devices are made aware of this Urgent Medical Device Correc
Quantity4772 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190910PrisMax System Version 2
Baxter Healthcare Corp, Jul 25, 2019