ClearanceBrief
Recall Z-0269-2026 ยท IZL System, X-Ray, Mobile

SEDECAL SA

Model Number SM-40HF-B-D-C; 40KW 55C, Mobile X-ray system

DateOct 23, 2025 (initiated Jun 10, 2025)
StatusOpen, Classified
ReasonSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
Root causeDevice Design
ActionSedecal emailed their distributors on 06/10/2025 to remind them the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning. on 09/19/2025 the firm issued an Important Safety Notice to its direct consignees via email. This notice initiates Phase 2 of the recall strategy expanding the scope and providing information on stickers to cover gaps around the fine positioning control buttons and a DMC software update that stops unit movement if triggered by a short cicui
Quantity1 unit
DistributionUS: CA, IL, and NJ
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090322EASY MOVING PLUS, SM-XXHF-YY
Sedecal., Sa., Mar 17, 2009