Recall Z-2230-2025 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Siemens Healthcare Diagnostics Inc

RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.

DateJul 31, 2025 (initiated Jun 26, 2025)
StatusOpen, Classified
ReasonSiemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not cau
Root causeUnder Investigation by firm
ActionSiemens Healthcare sent an Urgent Medical Device Correction (UMDC) letter to consignees (via FedEx in the US) on around 06/26/2025. The communication letter informs consignees of the issue and provided actions for consignees to take. Consignees were instructed to review the letter with the Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable, replace the both the measurement cartridge and the wash/waste cartridge if still using a wash/waste cartridge from lot WW/08925 with RAPIDPoint systems, discontinue use of the wash/
Quantity1,477 units
Distributionworldwide distribution - US Nationwide and the countries of Canada, Chile, Japan, Paraguay.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002738BAYER RAPIDPOINT 400 SYSTEM
Bayer Corp., Dec 19, 2000