Recall Z-2098-2025 ยท KPS System, Tomography, Computed, Emission

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING

Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System

DateJul 8, 2025 (initiated Jun 20, 2025)
StatusOpen, Classified
ReasonUnintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
Root causeSoftware design
ActionFirm notified customers via Urgent Medical Device Correction letter dated June 20, 2025. Customers were instructed to check gantry software version. Instructions to do so are included in the letter. If system has the correct gantry software version, customers may continue to use their system--V.25BK for Infinia Running on Windows2000 (Win2K); C.15BK or C.1523 for Infinia on Windows XP; V.60BK on VG (all operating systems). If the software version is not the correct version, customers should discontinue use of system and contact GE HealthCare Service or local Service Representative. GE Healt
Quantity21
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953801VARICAM
Elscint, Inc., Nov 22, 1995