Recall Z-2221-2025 ยท OBJ Catheter, Ultrasound, Intravascular
Volcano Corp
Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 86700J used with Instructions for use.
| Date | Jul 30, 2025 (initiated Jun 13, 2025) |
| Status | Open, Classified |
| Reason | Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating instructions for use and adding: "Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire" |
| Action | On 6/13/2025, correction notices were mailed to customers who were asked to do the following: As per standard practice, take precautions when advancing or removing a catheter in complex vessel anatomies. Do not force a catheter into a narrow vessel or tight stenosis. Be aware that vessel calcification, tortuosity, and untreated vessel spasm are key components of complex vascular anatomy. If the initial guidewire faces resistance or requires further manipulation, consider this to be an indicator of complex patient anatomy. The Instructions for Use currently contains the following related preca |
| Quantity | 233,817 |
| Distribution | Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries |
| FDA record | Open on fda.gov |