ClearanceBrief
Recall Z-2310-2025 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Medical Systems, LLC

GE SIGNA Artist Evo, Nuclear Magnetic Resonance Imaging System

DateAug 13, 2025 (initiated Jun 20, 2025)
StatusOpen, Classified
ReasonGE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned mainten
Root causeEquipment maintenance
ActionGE HealthCare issued an URGENT CORRECTION NOTICE to it consignees on 06/20/2025 via traceable means. The notice explained the safety issue and requested the following actions be taken: "You can continue to use your MR system by following the instructions below: In the undocked position, lock the table by pressing the pedal in the front of the table. Check to see that all casters are locked and that the table is secure. Contact your GE HealthCare representative for assistance if any of the casters do not lock. Before each patient transfer verify that the caster locks are engaged. If you no
Quantity138 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213603SIGNA Artist Evo
Ge Healthcare (Tianjin) Company Limited, Feb 11, 2022