Recall Z-2213-2025 · LNH System, Nuclear Magnetic Resonance Imaging
Philips North America
MR 5300 Model Numbers (REF): (1) 782110, (2) 782135 (China ONLY), (3) 782152;
| Date | Jul 30, 2025 (initiated Jun 30, 2025) |
| Status | Open, Classified |
| Reason | The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire. |
| Root cause | Device Design |
| Action | On June 30, 2025, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. 4. Actions that should be taken by the customer / user in order to prevent risks for patients or users 4.1. When using any affected system identified according to the information in section 3, follow the instructions below. 4.2. Pay particular attention if a scan interruption occurs and a scan abort symbol is encountered, which may appear in the user interface (UI) with the symbols shown in Figure 4 below: Figure 4. Scan Abort Symbols 4.3. When scan aborts occur the symbol is shown and applicable messages are |
| Quantity | 551 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cu |
| FDA record | Open on fda.gov |