Recall Z-2112-2025 ยท REA High-Power Laser Light Show Projector

Beyond Laser Systems, LLC

Laser light show projectors manufactured by BLS, Model CLUB 1000/2000/4000.

DateJul 16, 2025 (initiated Jun 30, 2025)
StatusOpen, Classified
ReasonThe remote interlock connector, which is required by 21 CFR 1040.10(f)(3), in the remote stop box was not installed the connector or was mechanically installed but not wired. The labels placed on the projectors do not meet requirements specified in 21 CFR 1010.4 (Variances), 21 CFR 1010.2 (Certification), 21 CFR 1010.3 (Identification) and 21 CFR 1040.10 (g)(Labelling requirements). The identification label does not include the date of manufacture as required by 21 CFR 1010.3. The corrective action plan covers three families of laser light show of projectors: DIODE. CLUB and PURE with three d
Root causeRadiation Control for Health and Safety Act
ActionBeyond Laser Systems (BLS) mail the CAP letters to all affected customers. BLS will complete the procedure and ship the laser projector back to the customer within 14-21 days of receiving the units in our facility. BLS will provide a UPS shipping label to the customer for shipping to our facility and cover the expense of shipping the units back to the customer via UPS. If you have any questions, call 512-220-9339 ext 302.
Quantity12
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov