Recall Z-2196-2025 ยท EGJ Device, Iontophoresis, Other Uses

LTS Therapy Systems, LLC

IontoPatch STAT, On-the-Go Patch Therapy, Model/Catalog Number D-0062-080; IontoPatch is an Iontophoresis transdermal patch.

DateJul 29, 2025 (initiated May 22, 2025)
StatusOpen, Classified
ReasonThe nonwoven pads in the device do not absorb the saline solution or any other liquid intended to be used with the device.
Root causeUnder Investigation by firm
ActionLTS issued an URGENT: Medical Device Removal, Initial Notification to its consignees on 05/22/2025 via telephone and email. The notice explained the issue, the risk to health, and requested the following: Actions to be taken by the distributor: 1. Please stop all distribution of IontoPatch STAT Lot # s 4283-1, 4283-2, 4283-3, 4283-4 and IontoPatch 80 Lot # s 4303-1, 4303-2, 4303-3. 2. Contact your clients to which these product lot # s have been distributed and instruct them to cease all sales, distribution and/or usage. Further Actions: 1. This initial notification is to stop further di
Quantity80928 units
DistributionUS Nationwide distribution in the states of Indiana, Tennessee, New York, Minnesota, New Jersey, Ohio, Kentucky, Illinois, and South Dakota.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K992708IONTOPATCH
Birch Point Medical, Inc., Feb 1, 2000