ClearanceBrief
Recall Z-2558-2025 ยท NHL Stimulator, Electrical, Implanted, For Parkinsonian Symptoms

Boston Scientific Neuromodulation Corporation

Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09

DateSep 12, 2025 (initiated Jul 8, 2025)
StatusOpen, Classified
ReasonWire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.
Root causeLabeling design
ActionOn 7/8/2025, correction notices were mailed in the U.S. and mailed, emailed, or hand delivered outside the U.S. Healthcare Professionals were asked to do the following: 1) Review the Surgical Implant Manual updates. 2) If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them. 3) A patient letter is enclosed with this communication, which can be shared with your patient and/or included within the patient's medical record. 4) Monitor per relevant IFU recommendations for any clinical observations of high mon
Quantity25,260
DistributionWorldwide - US Nationwide distribution including in the states of MN, FL, CO, MD, NY, NE, PA, WI, AZ, TN, MO, TX, MI, MT, MA, WV, AL, VA, NC, CA, OR, WA, LA, OH, NJ, NH, SD, GA, CT, NV, UT, IN, ID, IL, ME, SC, DC, OK, KY, AK, HI, RI, AR, KS, NM and the countries of Argentina, Brazil, Chile, Ecuador,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.