Recall Z-2237-2025 ยท GKZ Counter, Differential Cell

Becton, Dickinson and Company, BD Bio Sciences

BD Trucount Controls, 30 Test - REF: 664343, RxOnly, IVD

DateAug 1, 2025 (initiated Jul 8, 2025)
StatusOpen, Classified
ReasonControl kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.
Root causeLabeling design
ActionOn July 8, 2025, BD Biosciences issued a "Urgent: Medical Device Recall Correction" via mail. BD ask consignees to take the following actions: 1. Ensure the contents of this product notification are read and understood. 2. Share and post this customer letter with the users of this application at your facility and forward to any customers to whom you may have distributed the product to ensure awareness. 3. It is recommended that this notice be posted where these products are stored. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not
Quantity10 kits
DistributionWorldwide - US Nationwide distribution in the states of Tennessee, New York, Massachusetts and the countries of Argentina, Belgium, Botswana, Brazil, Canada, France, Germany, Lithuania, Peru, Philippines, United Kingdom.