ClearanceBrief
Recall Z-2020-2025 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

interventional fluoroscopic x-ray system

DateJun 27, 2025 (initiated May 15, 2025)
StatusOpen, Classified
ReasonIn very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
Root causeRadiation Control for Health and Safety Act
ActionSiemens Healthineers sent customers a Customer Advisory Notice dated May 15, 2025 informing customers of a potential issue with ARTIS pheno and ARTIS icono systems with the following model numbers: ARTIS pheno 10849000, ARTIS icono biplane 11327600, ARTIS icono ceiling 11328100, ARTIS icono floor 11327700; due to in very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose and appropriate software update will be implemented, based on
DistributionWorldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K241572ARTIS icono (VE40A) (floor); ARTIS icono (VE40A) (biplane); ARTIS icono (VE40A) (ceiling); ARTIS pheno (VE40A)
Siemens Medical Solutions USA, Inc., Oct 22, 2024