Recall Z-2152-2025 ยท CGF Radioimmunoassay, Estrone

Beckman Coulter Inc.

Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.

DateJul 21, 2025 (initiated Jun 12, 2025)
StatusOpen, Classified
ReasonA specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for the impacted Estrone RIA kits.
Root causeNonconforming Material/Component
ActionOn June 12, 2025, Beckman Coulter issued a "Urgent: Medical Device Recall" notification via Email. On July 8, 2025, Beckman Coulter issued a updated notification to provide additional information on the health risk and actions consignees should take. Beckman Coulter ask consignees to take the following actions: 1. Beckman Coulter recommends sharing the content of this letter with your laboratory and/or medical director regarding the need to review previous patient test results. 2. Please discard any remaining inventory of effective lots per your laboratory protocols and ask for replacement.
Quantity496 units
DistributionUS: North Carolina OUS: Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, Netherlands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K935013DSL ESTRONE (DSL 8700)
Diagnostic Systems Laboratories, Inc., Jun 16, 1994