Recall Z-2323-2025 ยท KFF Assay, Heparin

Instrumentation Laboratory

HemosIL LMW Heparin Controls; Part Number: 0020300200;

DateAug 13, 2025 (initiated Feb 3, 2025)
StatusOpen, Classified
ReasonMultiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL Liquid Anti-Xa and HemosIL Heparin Calibrator. In some complaints, failure to pass QC resulted in patient testing delays.
Root causeProcess control
ActionOn February 3, 2025, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Requested Customer Actions Please take the following actions: " Follow the instructions in the original Urgent Medical Device Correction letter to update the ACL TOP test parameters with the corrected value assignments for HemosIL Heparin Calibrators (Part No. 0020300600). " For HemosIL UF Heparin Controls (Part No. 0020300300), continue to use the current target values and acceptance ranges as labeled. " For HemosIL LMW Heparin Controls (Part No. 0020300200), use the reassigned target values and acceptance
Quantity41,824 units
DistributionWorldwide - US Nationwide and the countries of Algeria, Andorra, Argentina, Australia, Austria, Belarus, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, Germany, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, J
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213464HemosIL Liquid Anti-Xa
Instrumentation Laboratory CO, Oct 4, 2022
K980242IL TEST HEPARIN
Instrumentation Laboratory CO, Mar 19, 1998