Recall Z-2164-2025 ยท OZD Temporary Non-Roller Type Left Heart Support Blood Pump

Abiomed, Inc.

Automated Impella Controller (AIC), used in left heart support blood pump, including the following: 1. Impella Controller, Packaged, AU; Product Code: 0042-0000-AU. 2. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 3. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 4. Impella Controller, Packaged, EU, Loaner; Product Code: 0042-0000-EU-L. 5. Impella Controller,

DateJul 26, 2025 (initiated Jun 23, 2025)
StatusOpen, Classified
ReasonA potential issue with the Automated Impella Controller (AIC) not detecting an Impella pump when it is connected
Root causeDevice Design
ActionInitially Abiomed posted their recall notification on their website, HeartRecover.com on 06/23/2025. Consignees were also notified on 06/25/2025 via letter. Consignees were advised that devices were not being removed. They were advised to have a back-up Automated Impella Controller (AIC) available in the event of a device failure and were provided work around instruction for console-to-console transfer and case starts. Additionally, consignees were instructed to forward the notification to all pertinent personnel and customers if further distributed and complete and return the business respon
Quantity11,031 units
DistributionWorldwide distribution. US Nationwide, Australia, Austria, Belgium, Brunei Darussalam, Canada, Colombia, Croatia, Czechia, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.