Recall Z-2134-2025 ยท OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Azurion 7 M20. Fluoroscopic X-Ray System.

DateJul 18, 2025 (initiated Jun 4, 2025)
StatusOpen, Classified
ReasonPotential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose.
Root causeDevice Design
ActionA "Device URGENT Medical Device Correction" notification dated 6/03/2025 was mailed to consignees. Consignees are instructed to circulate the provided notification to all system users and keep a copy of the recall notification with affected devices to maintain awareness. If you notice unusual noise during C-arm roll motion, or if the C-arm roll motion is disabled and the user message Motorized movement is not available appears, immediately contact a Philips representative. Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your syst
Quantity78 units
DistributionWorldwide distribution: US (nationwide) to states of: CA, CO, DC, DE, FL, IL, IN, KY, MA, MD, MN, MO, MT, NJ, NV, NY, OH, SC, TX, WA; and OUS (Foreign) to countries of: Argentina (AR), Australia (AU), Canada (CA), Chile (CL), Germany (DE), Dominican Republic (DO), Egypt (EG), ES, France (FR), United
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163715Allura Xper R9
Philips Medical Systems Nederland B.V., Jan 26, 2017
K172822Azurion series R1.2
Philips Medical Systems Nederland B.V., Nov 22, 2017
K181830Azurion R2.0
Philips Medical Systems Nederland B.V., Aug 2, 2018