Recall Z-2570-2025 ยท MUJ System, Planning, Radiation Therapy Treatment

Philips Medical Systems (Cleveland) Inc

Pinnacle Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model 870258 UDI code: (01)00884838103566(10)18.0.5.22236 Pinnacle TumorLOC 16.2 Model 870226 UDI code: (01)00884838091122(10)16.2.0.60010 Pinnacle TumorLOC 16.2.1 Model 870227 UDI code: (01)00884838091122(10)16.2.1.60098 Pinnacle TumorLOC 16.0.2 Model 870218 UDI code: (01)00884838085602(10)16.0.2.60004 Pinn

DateSep 16, 2025 (initiated Aug 5, 2025)
StatusOpen, Classified
ReasonDue to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.
Root causeSoftware design
ActionOn 08/05/2025, the firm sent via FedEx an "Urgent Medical Device Correction" Letter informing customers of a software issue affecting the Pinnacle3 Radiation Therapy Planning System performing Advanced Region of Interest (ROI) Expansion and Contraction that results in a potential shift of the intended ROI. The software assumes the patient is always in the Head First Supine (HFS) position when transposing the Advanced Expansion to alternate positions. Customers are instructed to: -When using the Advanced ROI Expansion and Contraction feature in Pinnacle3, only use the Head First Supine (HFS)
Quantity810 (350 U.S. and 460 O.U.S.)
DistributionU.S Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041577PINNACLE3 RADIATION THERAPY PLANNING SYSTEM
Adac Laboratories, Jun 25, 2004
K090808PINNACLE3 SMARTARC
Philips Medical Systems, Inc., Apr 3, 2009
K170086Pinnacle Radiation Therapy Planning System
Philips Medical Systems (Cleveland), Inc., Feb 9, 2017