ClearanceBrief
Recall Z-1981-2025 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Siemens Medical Solutions USA, Inc

UROSKOP Omnia. Model Number: 10094910

DateJun 17, 2025 (initiated Jun 4, 2025)
StatusOpen, Classified
ReasonA resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.
Root causeDevice Design
ActionSiemens notified consignees on about 06/04/2025 via letter. Consignees were instructed that they can continue to use their system when paying attention for signs of burning smell and/or smoke coming from the device base. In this case, the system must be immediately disconnected from power supply via room emergency on/off switch and the Siemens Healthineers Service Organization must be informed. Siemens Healthineers is preparing an on-site field safety corrective action to replace the frequency inverter by another type that cannot cause the described problem. The field safety corrective acti
Quantity95 units
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101491UROSKOP OMNIA
Siemens Medical Solutions USA, Inc., Jun 30, 2010
K173639Luminos dRF Max, Luminos Agile Max, Uroskop Omnia Max, Multitom Rax
Siemens Medi Cal Solutions, Inc., Apr 2, 2018