Recall Z-2104-2025 ยท DXF Catheter, Septostomy

Boston Scientific Corporation

VersaCross RF Wire (VXW). Indicated for creation of an atrial septal defect in the heart. Product ID VXW0002 Device approved and commercialized in Japan only.

DateJul 10, 2025 (initiated Jun 10, 2025)
StatusOpen, Classified
ReasonPotential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
Root causePackaging change control
ActionOn June 10, 2025, the firm began notifying consignees of the field action via Urgent Medical Device Removal letters. Customers are instructed to not use affected product. They should remove affected units from their inventory and segregate the units in a secure place until they can be returned to Boston Scientific.
Quantity14,724
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.