ClearanceBrief
Recall Z-2092-2025 ยท EBI Resin, Denture, Relining, Repairing, Rebasing

STRATASYS LTD

TrueDent Magenta, TDM130, Model No. OBJ-09166 Light-curable methacrylate-based resin that enables fabrication of dental appliances

DateJul 7, 2025 (initiated Jan 19, 2024)
StatusCompleted
ReasonCustomers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.
Root causeSoftware design
ActionThe firm initiated the recall to the consignee level on January 15, 2024. Customers were asked to coordinate a cartridge replacement. The reporting of this recall was done by the firm retroactively; the issue was resolved in January 2024.
Quantity94
DistributionUS Nationwide distribution including in the states of Arizona, Arkansas, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Maryland, Massachusetts, Michigan, Nebraska, New York, North Carolina, North Dakota, Oklahoma, Oregon, Utah, Wisconsin.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K220771Stratasys TrueDent
Stratasys, Ltd., Jul 7, 2022