Recall Z-2103-2025 ยท DXF Catheter, Septostomy

Boston Scientific Corporation

VersaCross Access Solution (VXSK), Product Description and ID: VXSK (180J RF WIRE 63CM/45DEG/D0) VXSK0101 VXSK (180J RF WIRE 63CM/45DEG/D1) VXSK0102 VXSK (180J RF WIRE 63CM/55DEG/D0) VXSK0103 VXSK (180P RF WIRE 63CM/45DEG/D0) VXSK0121 VXSK (180P RF WIRE 63CM/45DEG/D1) VXSK0122 VXSK (180P RF WIRE 63CM/55DEG/D0) VXSK0123 VXSK (180P RF WIRE 63CM/55DEG/D1) VXSK0124 VXSK (180P RF WIRE 63CM/90DEG/D0) V

DateJul 10, 2025 (initiated Jun 10, 2025)
StatusOpen, Classified
ReasonPotential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
Root causePackaging change control
ActionOn June 10, 2025, the firm began notifying consignees of the field action via Urgent Medical Device Removal letters. Customers are instructed to not use affected product. They should remove affected units from their inventory and segregate the units in a secure place until they can be returned to Boston Scientific.
Quantity14,724
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150709ProTrack RF Anchor Wire
Baylis Medical Company, Inc., Jun 17, 2015
K183655VersaCross Transseptal Sheath
Baylis Medical Company, Inc., May 20, 2019