Recall Z-2014-2025 ยท CBK Ventilator, Continuous, Facility Use

Maquet Critical Care AB

Servo-n Ventilator System. Model Number: 6694800.

DateJun 27, 2025 (initiated May 30, 2025)
StatusOpen, Classified
ReasonPotential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.
Root causeUnder Investigation by firm
ActionMaquet Critical Care AB, a subsidiary of Getinge, notified consignees on about 05/30/2025 via FedEx letter. Consignees were instructed to configure all affected Servo ventilators with "Extended leakage test" set to "Off", per chapter 8.2.4 Configuration in the User's manuals and all users are recommended to follow the updated instructions for use of the User's manuals chapter 9.5.1 Patient circuit test, which is provided. Consignees are instructed to ensure all devices in inventory have been properly addressed as described above, to ensure the message is forwarded to all affected personnel an
Quantity7,688 units
DistributionUS Nationwide distribution and Puerto Rico. No international distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201874Servo-u Ventilator System 4.1, Servo-n Ventilator System 4.1, Servo-u MR Ventilator System 4.1
Maquet Critical Care AB, Apr 20, 2021