Recall Z-2058-2025 ยท ITX Transducer, Ultrasonic, Diagnostic

Philips Ultrasound, Inc

Philips S5-1 Transducer.

DateJul 2, 2025 (initiated May 28, 2025)
StatusOpen, Classified
ReasonUltrasound transducer devices were refurbished beyond their useful life.
Root causeProcess control
ActionAn URGENT Medical Device Recall notification dated 5/28/25 was mailed to consignees. The notification informs consignees that Philips Ultrasound will be replacing transducers that were refurbished beyond their useful life with a new transducer. Consignees may continue to use affected devices while awaiting replacement; consignees should use devices in accordance with their intended use and per the instructions provided int he User Manual. Users should perform visual inspection and electrical safety testing as indicated in the User Manual; if a device does not pass inspection, it should be remo
Quantity5,230 units
DistributionDomestic: Nationwide Distribution; Foreign: Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042540PHILIPS IU22 ULTRASOUND SYSTEM
Philips Ultrasound, Inc., Oct 4, 2004
K081802CX50 DIAGNOSTIC ULTRASOUND SYSTEM
Philips Ultrasound, Inc., Jul 11, 2008
K091804CX50 VERSION 2.0 DIAGNOSTIC ULTRASOUND SYSTEM AND TRANSDUCERS
Philips Ultrasound, Inc., Jul 17, 2009
K111513CX50 DIAGNOSTIC ULTRASOUND SYSTEM
Philips Ultrasound, Inc., Jun 24, 2011
K123754CX50 3.0 DIAGNOSTIC ULTRASOUND SYSTEM
Philips Ultrasound, Inc., Dec 21, 2012