ClearanceBrief
Recall Z-2158-2025 · CAH Filter, Bacterial, Breathing-Circuit

Draeger, Inc.

SafeStar 90 Plus Filter. Bidirectionally breathing system filter.

DateJul 31, 2025 (initiated Jun 16, 2025)
StatusOpen, Classified
ReasonReports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters.
Root causeProcess control
ActionAn initial Urgent Medical Device Recall notification dated June 2025 was mailed to consignees. The notification instructs consignees to not make therapeutic decisions based solely on individual measured values or parameters, regardless of the measuring point and the ventilator or anesthesia machine used. Consignees are to not connect the sample line to the filter sampling port; if an alternative sample port is not feasible, the firm requests that consignees use an alternative filter. All users of affected devices are to be notified of this recall notice and the notice forwarded to customers to
Quantity669,838 units
DistributionWorldwide - US Nationwide distribution including in the states of AK, AZ, CA, CO, CT, FL, ID, IL, IN, KAS, LA, MA, ME, MI, MN, MO, NC, NE, NM, NV, NY, OH, PA, SD, TN, TX, UT, WA, WI, WY and the countries of Angola, Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia-Herz., Brazil, Bulga
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K221836Filter CareStar Plus, Filter SafeStar Plus, Filter/HME TwinStar Plus
Drägerwerk AG & Co. KGaA, Dec 7, 2022