Recall Z-2034-2025 ยท JAK System, X-Ray, Tomography, Computed

Philips North America Llc

Spectral CT. Computed tomography X-ray system.

DateJun 26, 2025 (initiated May 29, 2025)
StatusOpen, Classified
ReasonDevices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional software issues that affect CT performance.
Root causeUnder Investigation by firm
ActionAn "URGENT Medical Device Correction" notice dated 5/21/25 was mailed to consignees. The notification instructs consignees on how to identify if their devices have affected software downloaded. Consignees may continue to use affected systems in accordance with the intended use and recommended actions provided in the recall notification in Appendix A. Appendix A summarizes the ten software issues and the actions recommended by the manufacturer that the consignee follow. Consignees are to complete and return the provided response form, forward the recall notification to users, and retain a copy
Quantity370 units
DistributionDomestic: AZ, CA, CO, FL, GA, HI, IN, KY, MA, MD, MN, NY, OH, OR, PA, TX, VT, WA, WV; Foreign: Australia, Austria, Belgium, China, Czech Republic, Denmark, Ecuador, France, Germany, Hong Kong, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Lithuania, Macao, Malaysia, Netherlands, New Zeal
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K203020Spectral CT
Philips Medical Systems Nederland B.V., Feb 26, 2021