ClearanceBrief
Recall Z-2032-2025 ยท FSA Lift, Patient, Non-Ac-Powered

Baxter Healthcare Corporation

Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.

DateJul 9, 2025 (initiated May 30, 2025)
StatusOpen, Classified
ReasonThe Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.
Root causeDevice Design
ActionBaxter issued an URGENT PRODUCT RECALL notice to is consignees on 05/30/2025 via USPS first class mail. The notice described the problem with the device, hazard, and requested the following: "Actions to be taken by customers 1. Please locate and stop using all Q-link 13 components on any of the products listed in the affected product table. 2. Contact Baxter Technical Support to discuss alternative options, if necessary, at 800-445-3720, option 2, between the hours of 8:00 am and 6:00 pm Eastern Time, Monday through Friday, or by email at HRC_Technical_Support@Baxter.com. 3. Please post t
Quantity9839 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.