Recall Z-1957-2025 ยท JTR Kit, Quality Control For Culture Media

Microbiologics Inc

KWIK-STICK, Quality control kit for culture media, Catalog Numbers 01065K, Oligella urethralis derived from ATCC 17960; 0621K, Staphylococcus aureus derived from NCTC 12493

DateJun 11, 2025 (initiated May 27, 2025)
StatusOpen, Classified
ReasonAffected products were packaged with the incorrect KWIK STIK assembly resulting in the product containing the wrong microorganism. 0621K was packaged 01065K swabs; 01065K was packaged with 0621K swabs. This may lead to delayed diagnosis.
Root causeMixed-up of materials/components
ActionOn May 27, 2025, the firm began notifying customers via Urgent Medical Device Recall letters. Customers were informed of the affected products containing the wrong microorganisms. Customers were instructed to review lab procedures to understand how this information affects their usage, and then to use or discard the product depending on the lab procedures and how the issue affects usage. Distributors of the affected product were instructed to contact end users who have received the affected items.
Quantity3 total
DistributionUS Nationwide distribution in the states of California and Pennsylvania and the country of New Zealand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K861022KWIK STIK(TM) LYFO(TM) DISK MICROORGANISMS
Micro-Bio-Logics, Apr 4, 1986