Recall Z-2027-2025 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Siemens Healthcare Diagnostics, Inc.

Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;

DateJun 25, 2025 (initiated May 22, 2025)
StatusOpen, Classified
ReasonContamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.
Root causeUnder Investigation by firm
ActionOn May 22, 2025 URGENT FIELD SAFETY NOTICE letters were sent to OUS customers and URGENT MEDICAL DEVICE CORRECTION letters were sent to US customers. Customer Actions Please discontinue use of and discard all inventory of the diluent lot number listed in the table above (Products Section). If lot number 0000195412 is currently on your analyzer, remove the bottle and replace with an alternate lot number. See Appendix for detailed instructions. Review your inventory of this product to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporti
Quantity1,625 units (1450 US, 175 OUS)
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Bahamas.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151767Trinidad CH system, TD-LYTE Integrated Multisensor (Na, K, CI), TD-LYTE IMT Standard A, TD-LYTE IMT Standard B + Salt Br
Siemens Healthcare Diagnostics, Inc., Dec 11, 2015