Recall Z-1997-2025 ยท QLG Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems

BIGFOOT BIOMEDICAL

Bigfoot Unity Diabetes Management System SW-300689-01 Bigfoot Inject v1 Pen Cap LCAP Firmware 1.3.0 Release Configuration, Rev H SW-300689-02 Bigfoot Inject v1 Pen Cap RCAP Firmware 1.3.0 Release Configuration, Rev H The Bigfoot Unity Diabetes Management System that include the Pen Caps (Rapid-Acting Insulin Pen Cap [RCAP] and Long-Acting Insulin Pen Cap [LCAP]) integrates continuous monitoring w

DateJun 20, 2025 (initiated May 21, 2025)
StatusOpen, Classified
ReasonDue to a software/firmware issue diabetes management system may provide a device error and cause the insulin pen to become unresponsive and unable to function. This error may pose a potential health risk for people living with diabetes as they may not have all information necessary to make a timely treatment decision.
Root causeSoftware design
ActionOn 05/21/2025, the firm sent via FedEx and USPS Priority Mail, "URGENT MEDICAL DEVICE RECALL" Letters to Customers and Prescribers informing them that the Bigfoot Unity Pen Caps, both Long-Acting and Rapid-Acting Pen Cap firmware version 1.3.0 may provide a screen with "Device Error - Contact Bigfoot Customer Care" and render the pen caps unresponsive and unable to function. Customer are instructed to: 1. If customers have the affected product identified above, we are asking you to return the product. We have sent you a return kit, which you should receive shortly. If the return kit does not
Quantity57 pen caps
DistributionU.S. Nationwide distribution in the states of AL, FL, GA, IA, IL, NY, PA, TX, and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K202145Bigfoot Unity Diabetes Management System
Bigfoot Biomedical, Inc., May 7, 2021