Recall Z-2079-2025 ยท GKZ Counter, Differential Cell

Becton, Dickinson and Company, BD Bio Sciences

REF: 666289 , BD FACS, 7-Color Setup Beads, 25 Runs, CE, IVD. Used in flow cytometry for fluorescence compensation.

DateJul 3, 2025 (initiated Jun 3, 2025)
StatusOpen, Classified
ReasonDue to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.
Root causeProcess control
ActionOn June 3, 2025, Becton, Dickinson issued a "Urgent: Medical Device Recall Correction" Notification to affected customers via mail and E-Mail. BD asked consignees to take the following action: 1: Immediately inspect your inventory for the specific catalog and lot numbers listed. It is recommended that the clinical laboratory staff review and segregate the affected lots. 2. Remove and discard prior affected label insert with incorrect spectral overlap factor values included on kit box prior to use. 3. The wrong SOF (Spectral Over Factors) values can be corrected by entering the right SOF val
Quantity278 units
DistributionUS: Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Iowa, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040026BD FACS 7-COLOR SETUP BEADS
Becton Dickinson Immunocytometry Systems, Sep 10, 2004