Recall Z-2023-2025 ยท MJA Susceptibility Test Powders, Antimycobacterial

Becton Dickinson & Co.

BD BACTEC MGIT 960 PZA Kit. IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to pyrazinamide (PZA).

DateJun 25, 2025 (initiated May 20, 2025)
StatusOpen, Classified
ReasonBD has received additional complaints of intermittent false resistance results for PZA during susceptibility testing of Mycobacterium tuberculosis isolates.
Root causeUnder Investigation by firm
Action***UPDATE 11/13/2025*** BD Issued an URGENT: Medical Device Correction letters were sent to customers. BD is pleased to announce that we have resumed production of a modified version of the BD BACTEC MGIT 960 PZA Kit which includes modified inoculation methods and reduced shelf-life. BD has conducted a thorough review of the performance from all supported inoculum sources using the synthetic raw material and the following change has been implemented for the BD BACTEC MGIT 960 PZA Kit. 1. Only inoculum prepared from MGIT tubes 3-5 days post-positivity can be currently supported. 2. Inoculum pre
Quantity12,256 eaches
DistributionWorldwide distribution - US Nationwide and the countries of AR, AU, BD, BE, BN, BR, CA, CN, CR, DO, GT, HK, HT, ID, IN, JP, KH, KR, MO, MX, MY, NP, NZ, PE, PH, PK, SG, SV, TH, TW, VN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021582BACTEC MGIT 960 PZA KIT
Becton, Dickinson & CO, Jul 13, 2002