ClearanceBrief
Recall Z-1946-2025 ยท QBJ Integrated Continuous Glucose Monitoring System, Factory Calibrated

Dexcom, Inc.

Dexcom One Continuous Glucose Monitoring System, SKUs(Part Number): STK-DO-006 (MT27409-1), STK-DO-013 (MT27409-1), STK-DO-103 (MT27409-2), STK-DO-109 (MT27409-2)

DateJun 16, 2025 (initiated May 12, 2025)
StatusOpen, Classified
ReasonDefective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.
Root causeNonconforming Material/Component
ActionOn May 12, 2025, Dexcom issued a "Urgent: Medical Device Correction" Notification to affected consignees via E-Mail. On June 9, 2025, Dexcom issued an Urgent: Medical Device Correction update Dexcom asked consignees to take the following actions: 1. Verify if your receiver is affected by this issue: Check the serial number for any receiver(s) you use: https://dexcom.com/en-us/checkreceiver 2. If your receiver is affected, contact Dexcom Technical Support at 1-844-478-1600 to coordinate return of the affected receiver and a no charge replacement. 3. Whether or not your receiver is affected b
Quantity38,127 units
Distributionworldwide distribution - US Nationwide and the countries of Andorra, Argentina Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Latvia, Liechtenstein, Lithuania, Luxembourg, M
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.