Recall Z-1994-2025 ยท DXJ Display, Cathode-Ray Tube, Medical

Barco N.V.

Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300, MNA-440 DEC -H V2/K9303311; MNA-620: MNA-620 DEC -H/K9306210, MNA-620 ENC -H/K9306220; and MNA-640: MNA-640 DEC -H/K9306410, MNA-640 ENC -H/K9306420; components of the NexxisOR system

DateJun 20, 2025 (initiated May 9, 2025)
StatusOpen, Classified
ReasonCathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might switch to failover and back multiple times, devices/sources/sinks might not be discovered/available, might lead to delayed/prolonged procedures
Root causeNonconforming Material/Component
ActionOn 5/9/2025, recall notices started being distributed to customers informing them of the following: 1) Assess & locate the non-installed affected units and contain them from further distribution. B) Assess & locate affected units that are installed at your customer side. Please make your customers aware of this notification. C) The impacted units will need to be sent back. Further information on the swaps will be provided in the coming days. D) Fill in and return the Acknowledgement Letter. If you don t have the Key Account Manager s contact details, or need other support, you can contact the
Quantity221
DistributionUS Nationwide distribution in the states of TX, OH.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173381Nexxis OR
Barco N.V., Feb 22, 2018