Recall Z-1869-2025 ยท FPA Set, Administration, Intravascular

Fresenius Kabi USA, LLC

IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.

DateJun 6, 2025 (initiated May 12, 2025)
StatusOpen, Classified
ReasonMis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.
Root causeUnder Investigation by firm
ActionConsignees received an "URGENT - VOLUNTARY RECALL" notification via email dated 5/11/25. This notification instructs consignees to discontinue use and distribution of affected devices immediately. Consignees are to check their inventory for affected devices and place them in quarantine until product can be returned for replacement. The provided recall notification is to be distributed to users of the affected devices. Consignees are instructed to return the completed Customer Reply Form via fax or email to facilitate return of affected devices for replacement. Consignees with any questions can
Quantity14,280 units (724 cases)
DistributionUS distribution: CA, CO, DE, GA, IL, MD, MI, MN, MS, NC, NJ, NV, OK, SC, TX, VA, WA & WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183311Ivenix Infusion System (IIS)
Ivenix, Inc., Jun 7, 2019