Recall Z-1996-2025 · JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter, Inc.

DxI 9000 Access Immunoassay Analyzer C11137

DateJun 20, 2025 (initiated May 7, 2025)
StatusOpen, Classified
ReasonAnalyzer has calibration issues where curves switched from passed to failed due to system errors during aspiration. This anomaly causes erroneous results and delays if unnoticed. RLU values may be misinterpreted as valid measurements by LIS.
Root causeSoftware design
ActionThe firm began notifying customers the week of May 7, 2025 via Urgent Medical Device Recall letters. Only analyzers running software version 1.20 are affected. Customers were provided with instructions on what to do if they observe a calibration that remains in the "In Progress" or "Presented" state. Beckman Coulter will release a service modification to change a configuration setting to prevent this issue. Your service representative will schedule the service modification installation when it is available. In the meantime, it is recommended that customers share the content of this letter wit
Quantity67 US; 178 OUS
DistributionWorldwide - US Nationwide distribution in the states of AR, CA, FL, GA, HI, ID, IL, IN, KY, LA, ME, MI, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, WA, WI and the countries of Algeria, Australia, Austria, Belgium, Brazil, C¿te d'Ivoire, Croatia, Czechia, France, Germany, Hungary, India, Ireland, Israel,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K221225Access TSH (3rd IS) Assay, DxI 9000 Access Immunoassay Analyzer
Beckman Coulter, Inc., Nov 10, 2022