ClearanceBrief
Recall Z-1967-2025 ยท FNM Mattress, Air Flotation, Alternating Pressure

Turncare, Inc

Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Software Version: 2.5.4-59 (Fix)

DateJun 12, 2025 (initiated Nov 1, 2023)
StatusOpen, Classified
ReasonA software issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy.
Root causeSoftware design
ActionBetween June 2023 and November 2023 software updates to resolve the issue were performed by TurnCare on affected units. The software update corrected an issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy. Successful update of the software was verified on each affected controller.
Quantity1081 units
DistributionUS Nationwide distribution in the states of Texas, Illinois, Pennsylvania, Florida, Connecticut
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.