Recall Z-1919-2025 ยท JOW Sleeve, Limb, Compressible

Cardinal Health 200, LLC

Kendall SCD Blended Comfort Sleeves Knee Length Small - For use with Kendall SCD SmartFlow SKU 84021B Kendall SCD Small Comfort sleeves when used with the Kendall SCD Smartflow Controller

DateJun 6, 2025 (initiated May 13, 2025)
StatusCompleted
ReasonIncorrect software logic may cause frequent E1 errors (system high pressure). Issue may cause delay in treatment or therapy.
Root causeSoftware change control
ActionOn May 13, 2025, the firm began to notify customers via Urgent Medical Device Product Correction letters. Customers were informed of the issue, which exclusively impacts use of the Comfort Small Knee Sleeve. Other sleeve sizes are unaffected and operate as intended with the SmartFlow Controller. The product remains operational if the Comfort Small Knee Sleeve and SmartFlow Controller configuration is not utilized during therapy. A software update on the SmartFlow Controller is necessary before using the SmartFlow controller with the Comfort Small Knee sleeve configuration. Customers should
Quantity6517 cases total
DistributionDomestic US distribution to AK, CA, FL, MA, TX. International distribution to Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K230028Kendall SCD SmartFlow Controller, Cardinal Health Element Sleeves, Kendall SCD Express Sleeves, Kendall SCD Express Foot
Cardinal Health200, LLC, Apr 3, 2023