Recall Z-2483-2025 ยท BSB Warmer, Blood, Non-Electromagnetic Radiation

3M Company

3M Ranger Blood/Fluid Warming High Flow Set Catalog Numbers 24355, 24370

DateSep 18, 2025 (initiated Apr 10, 2025)
StatusOpen, Classified
ReasonField action to clarify the flow rates and inlet fluid temperatures, along with location of the outlet temperature for Ranger Blood/Fluid Warming High Flow Sets.
Root causeDevice Design
ActionAn Urgent Medical Device Correction letter was sent to all their customers on 4/21/2025 to inform them that Solventum [formerly 3M Company] clarified flow rates related to the inlet fluid temperature. The device instructions for use (IFU) will be updated to reflect these changes. Customers are informed that there have been no changes to the design or composition of these products that prompted this notification. Customers are instructed to continue to use the products as indicated, considering the updated information provided in the letter. If the device is used at a flow rate outside of the u
Quantity530,470 units
DistributionWorldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Belgium, Bosnia and Herzegovina, Canada, Canary Islands, China, Costa Rica, Croatia, Czechia, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K973741BAIR HUGGER BLOOD/FLUID WARMER
Augustine Medical, Inc., Apr 30, 1998